Author
Dr.Sc Ed Connor

Published
20th April 2015

GC/MS analysis of a complex sample in DCM using hydrogen carrier gas

Summary

Labs worldwide have recently found that helium has been in short supply, leaving a number of labs without carrier gas. Added to this, helium prices have doubled over the past 10 years causing a number of labs to look into alternative carrier gases for GC, such as nitrogen and hydrogen. As well as lower price and unlimited availability, hydrogen has a number of potential advantages over helium, including potential for faster throughput, improved chromatography and better sample resolution.  

Introduction

This application note aims to demonstrate the injection of a complex, 76-component sample diluted in Dichloromethane (DCM) using hydrogen carrier gas. One potential problem highlighted with using hydrogen carrier gas is its potential for reaction with analytes and chlorinated solvents. This study aims to show that hydrogen carrier gas produced by a Peak hydrogen generator is suitable for GC/MS analyses with the Bruker Scion-SQ. Results show that hydrogen carrier gas can be used in conjunction with DCM as a solvent and with the correct injection configuration can give improved results compared with helium as carrier gas.

Experimental

0.5 µL of a 76-compound mixture (Restek Megamix cat. No. 31850) diluted in dichloromethane (DCM) was injected into a hot split/splitless inlet of a Bruker Scion SQ GC/MSD. To increase the amount of sample transferred onto the column and reduce the dwell time in the inlet, in order to reduce the potential for reaction between carrier gas and solvent, pulsed-split injection was performed for 20 seconds. Hydrogen served as the carrier gas at a rate of 1.0 mL/min.

GC and MSD conditions were as follows:

Would you like more information?    

 

MS:

Bruker SCION-SQ GC-MSD

MS source: 330 °C

Mass Range: m/z 45-500

2 min solvent delay, 120 ms scan time.

GC:

Column: BP-5MS column (20m x 0.18mm with 0.18 µm film thickness)

Oven: 45 °C (1 min hold) ramped at 30 °C/min to 310 °C (5 min. hold)

Injection: Pulsed-split injection (inlet temperature 290 °C , pressure 40 psi for 0.3 min, 70:1 split)

 

 

Carrier gas:

Peak Precision Trace Generator Hydrogen, 1 mL/min

 

Instrumentation

The sample was analysed using the Bruker Scion SQ GC/MSD, which has a lens-free ion source allowing users to easily switch between helium and hydrogen. The inert ion source requires less frequent cleaning and maintains consistent levels of sensitivity. 

Hydrogen was provided by the Peak Scientific Precision Trace generator, which produces hydrogen from deionised water using a proton exchange membrane (PEM) cell and purifies hydrogen to 99.9999% using a pressure swing adsorption (PSA) dryer system.  

Results

The results of this technical brief show that by using hydrogen as carrier gas, users can shorten run times compared to helium without compromising peak separation (Fig. 1.). The results also show that hydrogen can improve not only run time efficiency, but signal strength and resolution (Fig 1).

Here we show that by using pulsed-split injection with a high inlet pressure and high split ratio, a fast transfer of sample to the column can be achieved, avoiding the formation of HCl in the inlet through reaction of DCM and hydrogen carrier gas.

 GC-MS Analysis 1

Figure 1. Results showing faster elution of compounds and greater peak height when using Hydrogen as a carrier gas.

 

 GC-MS Analysis 2

Figure 2 shows that separation of Indeno(123-cd)pyrene and Dibenz(ah)anthracene was also improved by using hydrogen despite faster elution times with hydrogen as carrier gas.

 

Discussion

Hydrogen produced by the Peak Scientific Hydrogen Trace generator is a safe and suitable alternative to Helium for GC/MS carrier gas. With the installation of an in-oven Hydrogen leak detector, along with the failsafe mechanisms in place within the generator, lab users can work safely with Hydrogen.

Owing to its potential to form HCl which can cause damage to the inlet and column, chromatographers should be careful when using DCM as a solvent in conjunction with Hydrogen as carrier gas. This experiment shows that it is possible to use Hydrogen as carrier gas for standards and extracts that are required to be in DCM.

 

Ed Connor

Ed Connor DR.SC. is GC-MS Application Specialist, Peak Scientific, Inchinnan Business Park, Scotland, UK . Prior to joining Peak in February 2013, Ed completed his Dr.Sc. at ETH Zurich in Switzerland using GC-MS to look at herbivore induced plant volatiles and their interaction with beneficial insects. He then joined the University of Zurich where his work focused primarily on volatile collection methods and analyses using GC-MS and GC-FID . +44 141 812 8100, econnor@peakscientific.com

 

Notes

  1.        Find more information on the Bruker SCION-SQ™ GC-MS
  2.        Find more information on the Peak Precision Hydrogen Trace generator

Trademarks

SCION-SQ™ is a Trademark of Bruker Corporation.

Subscribe to our newsletter for regular product updates, news and insights.